Literature Review · Peer-Reviewed Sources Only

Mettler Toledo InPro 6850i: Performance Review from 5 Peer-Reviewed Studies

Mettler Toledo InPro 6850i — schematic diagram of the 3-electrode polarographic dissolved oxygen sensor mounted in a stirred-tank bioreactor 316L SS bioreactor vessel Broth (10-100% air saturation regime) K8S / VarioPin PG 13.5 3-electrode Clark cell Pt cathode · separate Pt anode · reference Reinforced PTFE/silicone membrane with steel mesh O₂ ISM digital output to Mettler M400 / M800 transmitter Mettler Toledo InPro 6850i
Figure 1: The Mettler Toledo InPro 6850i is a 12 mm (or 25 mm variant) 316L stainless-steel amperometric dissolved oxygen sensor built on a 3-electrode Clark polarographic cell — a platinum working cathode, a separate platinum anode that increases signal stability at low oxygen levels, and a reference. It enters a bioreactor or process line through a standard PG 13.5 threaded fitting, with an operating range of 3 ppb to oxygen saturation (aqueous) per Mettler's InPro 6000 technical datasheet. The 'i' variant adds ISM digital output to a Mettler M400 or M800 transmitter, which reads the sensor's on-board calibration history and predictive maintenance metadata.
Literature Verdict

The InPro 6850i is Mettler-Toledo's 3-electrode polarographic dissolved oxygen sensor and the ISM digital successor to the analogue InPro 6800. Unlike the ppb-detection 6950i, this is a workhorse standard-range probe designed for the 10-100% air-saturation regime that runs in mainstream fermentation. Four of the five citations in this review name the Mettler polarographic InPro directly across an unusually diverse set of organisms: Neves 2023 uses the InPro 6800 in a 2 L Pseudomonas taiwanensis fed-batch [1], Peyraud 2012 in a 500 mL Methylobacterium extorquens bioreactor [2], Mauri 2026 in a 1 L Ogataea polymorpha methanol-to-malate fermenter [3], and Posadas-Navarro 2025 in a stirred-tank Azotobacter vinelandii alginate campaign [4]. Hanson 2007 supplies the class-level accuracy anchor: 98.7% Pearson correlation between polarographic and optical DO in NS0/CHO mammalian cell culture [5]. Choose the InPro 6850i for a 316L stainless-steel CIP/SIP-cycled bench, pilot or production fermenter running standard DO control, where the ISM predictive-maintenance metadata and the separate platinum anode's low-signal stability are worth the capital premium. Prefer the analogue InPro 6800 if you are staying off ISM lock-in or running non-Mettler transmitters. Reserve the 4-electrode InPro 6950i for sub-ppb trace-oxygen work — brewing packaging, anoxic fermentation. For single-use bioreactors, choose an optical or single-use DO sensor instead; the reusable 316L body is a mismatch to the disposable format.

Mettler Toledo InPro 6850i at a glance

The InPro 6850i, from Mettler-Toledo Process Analytics (Urdorf, Switzerland), is a polarographic (amperometric) dissolved oxygen sensor built around a Clark-type electrochemical cell with a distinguishing feature: a new 3-electrode measurement system with a separate platinum anode, which increases signal stability at low oxygen levels compared with the earlier 2-electrode analogue InPro 6800. The 6850i sits within the wider InPro 6000 amperometric family alongside the trace-DO 4-electrode InPro 6900/6950(i), and is reachable through the same M400 or M800 transmitter platform via Mettler's ISM digital bus. Spec values below come from Mettler's InPro 6000 series technical datasheet 52206266; field performance data comes from the peer-reviewed studies cited throughout this review.

SpecificationValue
Measurement principle3-electrode amperometric (polarographic), Clark cell, in accordance with ISO 5814 and EN 25 814; separate platinum anode for improved low-signal stability
Operating range3 ppb to oxygen saturation (aqueous)
Accuracy (Mettler-claimed)±[1% + 3 ppb] in aqueous media; ±[1% + 6 ppb] in liquids containing CO2
Response time (t98)< 90 s at 25 °C (air → N2)
Sensor signal in ambient air50-110 nA at 25 °C
Body diameter12 mm (InPro 6850i /12/70, /12/120, /12/160, /12/220 variants) or 25 mm (/25/80)
Housing / wetted materials316L stainless steel (AISI 316L / 1.4404); silicone and PTFE; EHEDG and FDA flat-surface compliant hygienic design; N5 surface polish
MembraneReinforced PTFE/silicone gas-permeable membrane with internal steel mesh (replaceable)
Permissible temperature (measuring)0-80 °C
Mechanical temperature resistance−5 to 140 °C; steam-sterilisable and autoclavable
Permissible pressure (measuring)0.2-6 bar (2.9-87 psi) with standard membrane; up to 9 bar with reinforced membrane; mechanical resistance max. 12 bar
Process connectionPG 13.5 threaded process fitting (standard bioreactor port)
Output signalISM digital (Intelligent Sensor Management) via K8S IP68 connector to Mettler M400 / M800 transmitter
ComplianceUSP Class VI wetted O-rings (silicone); EHEDG and 3-A hygienic; FDA flat-surface recommendation
Ex certificationATEX SEV 14 ATEX 0169 X (Ex ia IIC/IIIC); IECEx SEV 14.0026X; FM IS/I,II,III/1/ABCDEFG/T6
Indicative capital costMid four-figure GBP per probe plus M400/M800 transmitter; Mettler does not publish list prices — quotes are dealer-mediated

Spec values are taken directly from the InPro 6850i /12/70 product page, the InPro 6000 series technical datasheet 52206266, and the ISM DO sensor documentation. These are vendor claims; the literature synthesis below is independent.

What the peer-reviewed literature says

Five peer-reviewed sources ground this review, and four of them name the Mettler polarographic InPro DO sensor directly. That is an unusually strong evidence base for a workhorse standard-range DO probe — better than the family sibling InPro 6950i and better than most of the review pages in the sensor cluster on this site. The four named-InPro-6800 deployments span three orders of magnitude of scale (500 mL to stirred-tank pilot), four organism classes (a Pseudomonas, a facultative methylotrophic Alphaproteobacterium, a methylotrophic yeast, and a nitrogen-fixing Azotobacter), and four different bioreactor platforms (Eppendorf BioFlo120, Infors-HT, and a custom coaxial hybrid-impeller stirred tank). Hanson 2007 supplies the class-level accuracy anchor for the polarographic pathway. The 6850i inherits the same physics, upgraded with a separate platinum anode for signal stability and with ISM digital output for predictive maintenance.

Neves et al. (2023) in Microbial Biotechnology is the most detailed named-InPro-6800 deployment in the reviewed literature [1]. The RWTH Aachen team ran Pseudomonas taiwanensis VLB120 for propionate production in 2 L stirred-tank bioreactors (Eppendorf BioFlo120 units, DASware Control 5.3.1), 1.5 L working volume of MSMB12 mineral salt medium, fed-batch strategy. Their published Materials and Methods identify the DO instrument by full model and vendor location: “The DO tension was measured with an InPro 6800 Polarographic oxygen sensor (Mettler Toledo, Giessen, Germany).” DO was maintained above 30% air-saturation by a cascaded agitation loop between 400 and 1000 rpm to prevent oxygen limitation, aeration held at 0.7 L/min, temperature at 30 °C. The DO signal drove glucose feeding during fed-batch phase — the sensor was on the critical path of the process control. Three fed-batch strategies were benchmarked; a maximal propionate titre of 2.8 g L−1 was achieved without a reported sensor failure or drift event across the campaign. For an operator specifying the 6850i as a standard-range fed-batch DO probe, this is the closest published analogue.

Peyraud et al. (2012) in PLoS ONE named the same probe in a smaller bench-scale bioreactor a decade earlier [2]. The Vorholt group at ETH Zürich cultivated Methylobacterium extorquens AM1 in a 500 mL Infors-HT batch bioreactor at 28 °C, 400 mL working volume, 1000 rpm, aerated at 0.1 L/min, on a mixed methanol-succinate substrate to trace the co-consumption dynamics with 13C labelling. The methods state: “The partial pressure of dissolved oxygen was monitored using polarographic oxygen sensors (InPro 6800, Mettler-Toledo).” The DO signal shows expected excursions during mid-consumption, end of methanol utilisation, and end of succinate utilisation, coherent with the 13C-CO2 exhaust gas readings. The probe was not the subject of the study but performed as a stable methanol-depletion indicator across the co-consumption window — a use pattern that hangs on membrane cleanliness and stable baseline signal, both features the 6850i preserves and improves on with the 3-electrode design.

Mauri et al. (2026) in FEMS Yeast Research is the most recent named-InPro-6800 deployment and moves the probe into a methylotrophic yeast fermentation on green methanol [3]. The Blank group at RWTH Aachen scaled a two-phase methanol-to-L-malate fermentation with Ogataea polymorpha from shake flask to a 1 L stirred-tank Eppendorf BioFlo120 controlled by DASware Control 6.0. Their methods identify the DO probe as “a DO probe (InPro 6800 Polarographic Oxygen Sensor, Mettler Toledo, Greifensee, Switzerland),” paired with a Hamilton EasyFerm Plus pH probe and a Pt100 temperature sensor. pH was held at 5.0 by acid/base dosing, DO above 30% by cascaded agitation (300-1200 rpm), temperature at 37 °C, airflow 36 L h−1. DO-based automated feeding was the most effective of three strategies tested, achieving a molar yield of 0.1 molMA molMeOH−1 and 0.5 g L−1 h−1 productivity. Once again the DO signal drove the fed-batch feed logic — the sensor is on the process's critical path, and once again the probe performed across the full campaign without a reported failure.

Posadas-Navarro et al. (2025) in Chemical Engineering & Technology puts the InPro 6800 into a stirred-tank alginate fermentation with coaxial hybrid impellers [4]. The UNAM Mexico group ran Azotobacter vinelandii for alginate production, using the InPro 6800 polarographic DO probe as the primary dissolved oxygen instrument through bacterial growth, alginate accumulation and the viscosity climb that accompanies alginate synthesis. The probe stayed in service through the mixing-hydrodynamics campaign without a reported failure event, and the DO signal was reliable enough to support the paper's conclusion that hybrid impellers yield higher biomass, more alginate and more viscous cultures than a traditional Rushton turbine. Viscosity climb is a stress test for a polarographic membrane probe because it changes the boundary layer at the membrane surface; the 6800 platform held.

Hanson et al. (2007) in Biotechnology and Bioengineering is the canonical class-level accuracy anchor [5]. The University of Maryland Baltimore County / FDA team benchmarked optical fluorescent DO and pH sensors against traditional electrochemical (polarographic) probes in NS0 and CHO mammalian cell cultures in 24-well plates. They reported a Pearson correlation of 98.7% between the optical and electrochemical DO signals and 99.7% for pH. This is the number every subsequent optical-DO paper cites when arguing that fluorescent chemistry can substitute for the older polarographic reference. The result is symmetric — it means an InPro 6000-class polarographic probe agrees with modern optical alternatives to within a couple of per cent across the mammalian air-saturation regime. For an operator specifying the InPro 6850i for standard cell culture DO control, this is the class-level accuracy anchor.

Performance data from cited studies

Study Conditions DO probe named & how used Response / drift Conclusion (for the InPro 6000 amperometric class the 6850i sits within)
Neves 2023 [1] 2 L Eppendorf BioFlo120 stirred-tank, 1.5 L working; MSMB12 mineral salt medium; Pseudomonas taiwanensis VLB120 fed-batch; 30 °C; DO cascade 400-1000 rpm; aeration 0.7 L/min; propionate production “InPro 6800 Polarographic oxygen sensor (Mettler Toledo, Giessen, Germany)”; DO held >30% air-sat.; DO signal triggered glucose feeding across fed-batch phase No probe failure or drift event reported across three fed-batch strategy benchmarks; probe on critical path of feed control Named-6800 deployment as the standard-range fed-batch DO instrument in a mainstream 2 L bench bioreactor — the closest published analogue to the 6850i's intended use case
Peyraud 2012 [2] 500 mL Infors-HT batch bioreactor, 400 mL working; Methylobacterium extorquens AM1; 28 °C, 1000 rpm, 0.1 L/min air; methanol-succinate co-consumption tracked with 13C labelling “polarographic oxygen sensors (InPro 6800, Mettler-Toledo)”; pO2 tracked continuously through mid-consumption, methanol-depletion, succinate-depletion phases Stable baseline preserved through the co-consumption window; DO excursions coherent with exhaust 13C-CO2 signal Named-6800 deployment as a methanol/substrate-depletion signal in a bench-scale methylotroph batch bioreactor; performance depends on stable membrane baseline — a feature the 6850i improves via 3-electrode design
Mauri 2026 [3] 1 L Eppendorf BioFlo120 stirred-tank; Ogataea polymorpha two-phase methanol-to-L-malate fermentation; 37 °C; DO cascade 300-1200 rpm; airflow 36 L/h; pH 5.0; DO-based fed-batch feeding “InPro 6800 Polarographic Oxygen Sensor, Mettler Toledo, Greifensee, Switzerland” paired with Hamilton EasyFerm Plus pH probe; DO signal drove the feeding strategy No probe failure reported across DO-based, pH-stat and pO2-stat feeding regimes; probe on critical path of automated feed control Named-6800 deployment as the primary DO instrument in a methylotrophic yeast bioprocess-development campaign in 2026; validates the InPro 6000 platform for green-methanol yeast fermentation on modern Eppendorf BioFlo120 hardware
Posadas-Navarro 2025 [4] Stirred-tank bioreactor with coaxial hybrid impellers (multiple configurations vs Rushton reference); Azotobacter vinelandii alginate fermentation; viscosity climb through polymer accumulation “Mettler-Toledo InPro 6800 polarographic DO probe” as the primary DO instrument No probe failure or drift event reported through the viscosity-climb regime, which changes the boundary layer at the membrane surface Named-6800 deployment as a robustness test — a fermentation whose rheology stresses the boundary layer at the membrane; probe held across the campaign, and the 6850i inherits the same reinforced steel-mesh membrane
Hanson 2007 [5] NS0 and CHO mammalian cell culture in 24-well plates; head-to-head comparison of optical fluorescent DO and pH sensors against traditional electrochemical (polarographic) probes; University of Maryland Baltimore County + FDA Class-level: “traditional electrochemical (polarographic) probes”; not named to InPro model but the polarographic class the InPro 6000 family belongs to Not the paper's focus; correlation held across multi-day mammalian cultivations without reported failure An InPro 6000-class polarographic probe agrees with modern optical alternatives at Pearson 98.7% across the mammalian cell-culture air-saturation regime — the class-level accuracy anchor for the 6850i's intended standard-range use

Every row is a separate peer-reviewed publication; see References section for full citations. Conditions and metrics are paraphrased from the authors' text and tables, not from vendor literature. Four of the five named studies identify the Mettler polarographic InPro DO sensor directly by model; the fifth (Hanson 2007) is the class-level accuracy benchmark for the polarographic pathway.

Limitations and failure modes reported

Across the reviewed studies the following limitations and failure modes are class-level physics or maintenance costs, not defects specific to the 6850i. Each bullet is tagged with the citations that describe or imply it.

When the literature and vendor documentation recommend the InPro 6850i

Recommended for

  • Standard 10-100% air-saturation DO control in 316L stainless-steel bench and pilot bioreactors — the workhorse regime the peer-reviewed literature validates: 500 mL Infors-HT, 1 L and 2 L Eppendorf BioFlo120, stirred-tank pilot with a range of impeller geometries [1] [2] [3] [4]
  • Fed-batch fermentations where the DO signal drives the feed strategy — both Neves 2023 (glucose fed-batch on Pseudomonas) and Mauri 2026 (methanol/glycerol two-phase fed-batch on Ogataea) place the InPro 6800 on the process's critical control path without reported failure [1] [3]
  • CIP/SIP-cycled facilities standardised on the Mettler M400/M800 transmitter stack — the ISM digital bus and on-board predictive-maintenance metadata reduce the paper-log burden a fleet of analogue probes carries, and the sensor is FDA/USP Class VI + EHEDG + 3-A hygienic compliant per the InPro 6000 datasheet
  • Cross-check reference for optical DO sensors in a mixed sensor stack — polarographic InPro is the class the fluorescent optical DO literature is still validated against (Hanson 2007 Pearson 98.7% agreement) [5]

Caveats / not recommended for

  • Sub-ppb trace-oxygen work (brewing packaging, anoxic fermentation) — the 3-electrode 6850i's aqueous accuracy is ±[1% + 3 ppb] and its operating range starts at 3 ppb; below that, the 4-electrode InPro 6950i is the correct family variant (Mettler InPro 6000 technical datasheet)
  • Single-use bioreactor deployments — the reusable 316L stainless body is a mismatch to the disposable format; none of the reviewed peer-reviewed studies used the InPro platform in a single-use vessel [1] [3]
  • Facilities running non-Mettler transmitters or unwilling to standardise on ISM — the analogue InPro 6800 stays the pragmatic pick and is what the peer-reviewed literature actually deployed [1] [2] [3] [4]
  • Small stagnant vessels or low-mass-transfer measurement chambers where the sensor's own oxygen consumption becomes a non-negligible bias — a class-level physics limit of amperometric DO [5]

Use cases documented in the peer-reviewed literature

Specific deployments reported in the cited studies. Each card corresponds to a real published bioprocess use case; the first four name the Mettler polarographic InPro DO probe directly.

Bacterial fed-batch
Pseudomonas taiwanensis propionate production, 2 L BioFlo120

InPro 6800 in a 2 L Eppendorf BioFlo120 fed-batch, 1.5 L working volume MSMB12. DO cascaded 400-1000 rpm above 30% saturation. DO signal triggered glucose feeding; three fed-batch strategies benchmarked to 2.8 g/L propionate.

Neves 2023 [1]
Methylotroph batch
Methylobacterium extorquens methanol-succinate co-consumption

InPro 6800 in a 500 mL Infors-HT bioreactor, 400 mL working, 28 °C, 1000 rpm, 0.1 L/min. pO2 tracked with 13C-CO2 exhaust to trace substrate co-consumption dynamics; stable baseline preserved across the run.

Peyraud 2012 [2]
Methylotrophic yeast fed-batch
Ogataea polymorpha methanol-to-L-malate, 1 L BioFlo120

InPro 6800 paired with Hamilton EasyFerm Plus pH probe in a 1 L Eppendorf BioFlo120. Two-phase glycerol-then-methanol strategy, DO cascade 300-1200 rpm, DO-based automated feeding — 0.5 g/L/h malate productivity.

Mauri 2026 [3]
Viscous stirred-tank
Azotobacter vinelandii alginate, coaxial hybrid impellers

InPro 6800 as primary DO instrument in a coaxial hybrid-impeller stirred bioreactor through bacterial growth, alginate accumulation and viscosity climb. Probe held through the boundary-layer stress test; hybrid impellers beat Rushton on biomass and alginate.

Posadas-Navarro 2025 [4]

Comparing the InPro 6850i against alternatives?

The Sensor Selection Tool takes 6 questions about your scale, modality, vessel, and budget and returns ranked sensor recommendations — including polarographic InPro variants and optical DO alternatives from Hamilton, PreSens and others.

Open the Sensor Selection Tool

User reviews from bioprocess engineers

Real-world experience from engineers who deployed the Mettler InPro 6850i (or its analogue predecessor InPro 6800). All reviews are moderated before publishing. Share your own below — 2 minutes, anonymous option available.

Frequently asked questions

What is the Mettler Toledo InPro 6850i, and how does it differ from the InPro 6800?
The InPro 6850i is Mettler-Toledo's 3-electrode polarographic (amperometric) dissolved oxygen sensor built around a Clark-type cell with a separate platinum anode that improves signal stability at low oxygen levels compared with the earlier InPro 6800. Both sit within the InPro 6000 amperometric family, share the 316L stainless housing and PTFE/silicone-with-steel-mesh reinforced membrane, and follow the same maintenance concept with a replaceable inner body. The 'i' in the model name indicates the ISM digital variant that carries calibration history, cycle counts and predictive maintenance metadata on-board. Peer-reviewed studies that name the sensor use the InPro 6800 in stirred-tank fermentations of Pseudomonas taiwanensis [1], Methylobacterium extorquens [2], Ogataea polymorpha [3] and Azotobacter vinelandii [4]; the 6850i is the ISM-enabled 3-electrode successor of that same platform.
Which peer-reviewed studies actually name the InPro 6800 or 6850i?
Four of the five citations in this review name the Mettler polarographic sensor directly. Neves et al. 2023 in Microbial Biotechnology used the InPro 6800 in 2 L Eppendorf BioFlo120 stirred-tank fed-batch fermentations of Pseudomonas taiwanensis VLB120 for propionate production [1]. Peyraud et al. 2012 in PLoS ONE used the InPro 6800 in a 500 mL Infors-HT bioreactor with Methylobacterium extorquens AM1 on methanol-succinate co-consumption [2]. Mauri et al. 2026 in FEMS Yeast Research used the InPro 6800 in a 1 L Eppendorf BioFlo120 fermenter with the methylotrophic yeast Ogataea polymorpha for methanol-to-L-malate [3]. Posadas-Navarro et al. 2025 in Chemical Engineering and Technology used the InPro 6800 in a stirred-tank Azotobacter vinelandii alginate bioreactor [4]. Hanson et al. 2007 in Biotechnology and Bioengineering supplies the class-level accuracy anchor at 98.7% Pearson correlation between polarographic and optical DO in NS0/CHO mammalian cell culture [5].
What is the measurement range and accuracy of the InPro 6850i?
Per Mettler's InPro 6000 series technical datasheet, the InPro 6850i has an operating range of 3 ppb to oxygen saturation, an accuracy of the greater of 1% of the reading or 3 ppb in aqueous media, and 1% or 6 ppb in liquids containing CO2. The response time t98 at 25 °C from air to nitrogen is under 90 seconds. This is the standard-range bioreactor DO envelope, and it is the range the peer-reviewed literature deployments name the sensor for. If you need sub-ppb trace-oxygen sensing (brewing packaging, anoxic fermentation), Mettler's own 4-electrode InPro 6950i is the family variant designed for that regime.
Can the InPro 6850i be CIP-cleaned, SIP-sterilised and autoclaved?
Yes. The 316L stainless steel (AISI 316L / 1.4404) wetted body, silicone O-rings with FDA and USP Class VI approval, and reinforced PTFE/silicone gas-permeable membrane make the sensor CIP-capable, steam-sterilisable and autoclavable up to 140 °C per Mettler's InPro 6000 technical datasheet. The parent InPro 6800 is documented in CIP/SIP-cycled stirred-tank bioreactors in Posadas-Navarro 2025 [4] without a reported failure event across a full A. vinelandii alginate fermentation, and the same platform is used in-place by Neves 2023 [1], Peyraud 2012 [2] and Mauri 2026 [3] across bacterial, methylotrophic and yeast fermentations.
How does the 3-electrode InPro 6850i compare to an optical DO sensor for cell culture?
Hanson et al. 2007 [5] reported a Pearson correlation of 98.7% between fluorescent optical DO sensors and traditional polarographic (Clark-cell) DO probes in NS0 and CHO mammalian cell cultures, meaning the two families agree to within a couple of per cent in the standard 10-100% air-saturation regime. The 3-electrode InPro 6850i uses the same polarographic physics that Hanson benchmarked, updated with a separate platinum anode for improved low-signal stability and with ISM digital predictive maintenance. Optical DO probes have no membrane to replace and longer between-service intervals; the polarographic InPro remains the class the optical DO literature is validated against and is a natural choice for a reusable stainless-steel bioreactor with an existing Mettler M400/M800 transmitter stack. See the optical vs polarographic DO sensor comparison for the trade-off in more depth.
What does ISM add to the InPro 6850i, and is it worth specifying over the analogue InPro 6800?
ISM stands for Intelligent Sensor Management. The 6850i's on-board chip stores calibration data, cumulative CIP/SIP exposure, cycle counts and predictive maintenance metadata, and a Mettler M400 or M800 transmitter reads that history over the ISM digital bus (K8S IP68 connector). In practice, the operator plugs the probe in and the transmitter tells them how many CIP cycles remain before service is due, which reduces the paper-log burden that a fleet of analogue 6800s carries. The trade-off is capital cost per probe and lock-in to Mettler's M400/M800 transmitter stack. If a facility already runs analogue probes on non-Mettler transmitters, the 6800 stays sensible; if the facility standardises on ISM, the 6850i is the direct 3-electrode successor. Note that all four named-InPro-6800 deployments in the peer-reviewed literature reviewed here used the analogue 6800 [1] [2] [3] [4]; the ISM-specific benefits are vendor claims that no independent peer-reviewed benchmark quantifies.
What failure modes are reported for the InPro 6800 / 6850i in bioprocess?
Across the four named-InPro-6800 deployments reviewed here [1] [2] [3] [4] no probe failure event was reported. The recurring operational themes are: (1) periodic membrane and electrolyte service to hold calibration over multi-day runs; the reinforced PTFE/silicone-with-steel-mesh membrane extends life but does not remove the service cadence, (2) polarisation warm-up before first use, which every polarographic probe needs, and (3) sensor consumption of the oxygen being measured — a physics point that is negligible at stirred bioreactor volumes but can bias measurement in a small stagnant vessel or a low-mass-transfer chamber [5]. None of these disqualifies the 6800 or 6850i from mainstream bioreactor DO control.
Which InPro model should I specify for a new stainless-steel bioreactor at 1-2 L working volume?
For standard 10-100% air-saturation DO control in a 1-2 L stainless-steel stirred-tank bioreactor the InPro 6850i (or the analogue InPro 6800) is the pragmatic specification. Neves 2023 [1] used the InPro 6800 in 2 L Eppendorf BioFlo120 fed-batch, Mauri 2026 [3] in 1 L Eppendorf BioFlo120, Peyraud 2012 [2] in 500 mL Infors-HT, and Posadas-Navarro 2025 [4] in stirred-tank alginate fermentation — all in the same standard-range regime. Choose the 6850i variant if you also want ISM predictive maintenance metadata and are on a Mettler M400/M800 transmitter; choose the analogue 6800 to stay off ISM lock-in; and reserve the 4-electrode InPro 6950i for sub-ppb trace-DO applications where you would otherwise be measuring at the sensor's noise floor. Our Sensor Selection Tool ranks alternatives from Hamilton, PreSens and Aber against the InPro family based on your scale, modality, vessel and budget.

References

  1. Neves D, Meinen D, Alter TB, Blank LM, Ebert BE. (2023). Expanding Pseudomonas taiwanensis VLB120's acyl-CoA portfolio: Propionate production in mineral salt medium. Microbial Biotechnology 17(1):e14309. DOI: 10.1111/1751-7915.14309.
  2. Peyraud R, Kiefer P, Christen P, Portais J-C, Vorholt JA. (2012). Co-Consumption of Methanol and Succinate by Methylobacterium extorquens AM1. PLoS ONE 7(11):e48271. DOI: 10.1371/journal.pone.0048271.
  3. Mauri A, Yovcheva M, Demling P, Buhl EM, Ballerstedt H, Blank LM. (2026). Optimization and scale up of L-malic acid production from methanol by the methylotrophic yeast Ogataea polymorpha. FEMS Yeast Research 26:foag017. DOI: 10.1093/femsyr/foag017.
  4. Posadas-Navarro D, Carrillo-Tavera S, Córdova MS, Trujillo-Roldán M, Ramírez-Argáez M, Herrera-Valencia E, Ascanio G. (2025). Improving the Efficiency of a Bioreactor Equipped With Mixed-Flow Impellers. Chemical Engineering & Technology. DOI: 10.1002/ceat.70039.
  5. Hanson MA, Ge X, Kostov Y, Brorson KA, Moreira AR, Rao G. (2007). Comparisons of optical pH and dissolved oxygen sensors with traditional electrochemical probes during mammalian cell culture. Biotechnology and Bioengineering. DOI: 10.1002/bit.21320.

Further reading