Miles & Misra vs Spread Plate vs Drop Plate: Which CFU Method Should You Use?
Pick the spread plate for release testing, Miles & Misra for scanning an unknown density, and the Herigstad optimised drop plate for high-throughput research. All three measure the same colony forming units per millilitre, but they trade plate count, media use, and regulatory acceptance. The spread plate is the pharmacopoeial default (Ph. Eur. 2.6.12, USP 61, FDA BAM Chapter 3); Miles & Misra covers six dilutions on one plate and ships with eighty-eight years of literature; the optimised drop plate (Herigstad et al. 2001) matches spread-plate precision at a fraction of the plastic.
Key differences at a glance
- Miles & Misra: 20 µL drops onto a sectored agar plate, six to eight dilutions on one plate, counts 2 to 20 colonies per drop.
- Spread plate: 100 to 200 µL spread across a full agar surface, one dilution per plate, 25 to 250 CFU per plate countable range (FDA BAM Chapter 3).
- Drop plate: 10 µL drops arranged in a grid of 20 to 25 replicates on one plate, one dilution per plate, 3 to 30 colonies per drop (Herigstad 2001).
- Best for regulatory release testing: Spread plate.
- Best for unknown starting density: Miles & Misra.
Side-by-side comparison
| Factor | Miles & Misra | Spread plate | Drop plate (Herigstad) |
|---|---|---|---|
| Inoculum volume | 20 µL per sector | 100 to 200 µL per plate | 10 µL per drop |
| Dilutions per plate | 6 to 8 (sectored) | 1 | 1 (replicated 20 to 25 times) |
| Countable range | 2 to 20 colonies per drop | 25 to 250 CFU per plate | 3 to 30 colonies per drop |
| Typical CV (replicated) | 15 to 20 percent | 10 to 15 percent | 10 to 15 percent |
| Plates per unknown sample | 1 plate scans 6 dilutions | 3 to 5 plates to bracket | 1 plate per dilution guess |
| Agar and plastic per count | Lowest (one plate) | Highest (one plate per dilution) | Low (one plate, many replicates) |
| Operator time per count | 8 to 12 min plating + 20 min counting | 5 min per plate plating + 10 min counting | 6 to 10 min plating + 15 min counting |
| Incubation time | 16 to 24 h (read small) | 18 to 48 h (standard) | 16 to 24 h (read small) |
| Regulatory acceptance | Research / in-house only | Pharmacopoeial (Ph. Eur., USP, BAM) | Research / high-throughput QC |
CV figures follow Hedges 2002. The spread-plate range of 25 to 250 CFU is the FDA BAM endorsement for aerobic plate counts.
Miles & Misra in detail
The surface viable count described by Miles, Misra, and Irwin in 1938 was designed to measure how fast whole blood kills Staphylococcus aureus. Their problem was logistical. A single bactericidal timepoint needs six serial dilutions, and six spread plates per timepoint was expensive in agar, plastic, and incubator space. The paper's answer was to divide one agar plate into six or eight equal sectors and pipette a 20 microlitre drop of each dilution onto its own sector. One plate, one timepoint, all six dilutions at once.
How it works
You prepare a 10-fold serial dilution series (typically 10⁻¹ through 10⁻⁶ or 10⁻⁷) in sterile diluent. A pre-dried agar plate is marked on its base into sectors. From the highest dilution first, 20 microlitre drops are pipetted onto the agar surface of each sector; the drops are allowed to soak in without spreading. The plate is incubated right-side-up at the species-appropriate temperature until drops produce discrete colonies, read while colonies are still small enough to be distinct (16 to 24 hours for E. coli at 37°C). The sector with 2 to 20 colonies per drop is the readable one, and the CFU per millilitre of the original sample is the colony count divided by the drop volume multiplied by the reciprocal of that sector's dilution factor. Three replicate drops per dilution is the minimum for a credible density estimate, which usually means two or three sectored plates total.
When Miles & Misra wins
Miles & Misra is the natural choice whenever you do not know the starting density within a factor of 10. A 10⁴-fold dilution range covered on one plate guarantees that at least one sector lands inside the countable window regardless of how far the starting density is from your guess. That is why environmental monitoring programmes in water, soil, and food pre-processing still use the sector layout. It is also the default for teaching microbiology, because every student's plate yields readable results on the first attempt without the operator knowing the sample strength. The method stays popular in quantitative microbiology textbooks as the "surface viable count" alongside pour and spread plating.
Spread plate in detail
The spread plate is the reference aerobic plate count method captured in FDA BAM Chapter 3, in European Pharmacopoeia 2.6.12, and in USP <61> for non-sterile pharmaceutical products. It is the default assay whenever a colony count ends up in a regulated document.
How it works
A 10-fold serial dilution series is prepared and 100 to 200 microlitres of one dilution is pipetted onto a pre-dried agar plate. The inoculum is spread across the entire agar surface with a bent glass rod, a disposable L-shaped spreader, or with 4 to 8 sterile glass beads that are rolled across the surface (the Copacabana method). The plate is inverted and incubated until countable colonies appear. The spread-plate countable range is 25 to 250 CFU per plate for most aerobic organisms (FDA BAM Chapter 3). CFU per millilitre is colony count divided by plated volume (in mL) multiplied by the reciprocal of the dilution factor. The method expects three to five plates covering adjacent dilutions so that at least one lands inside the 25 to 250 countable range.
When the spread plate wins
Use the spread plate whenever the result goes into a release certificate or a regulatory filing. It wins when the sample density is known within a factor of 10 so you can target the countable range with one or two dilutions. It also wins for colony-morphology work (presumptive identification, picking single colonies for downstream MALDI or sequencing) because each colony has a full field of agar around it and nothing is crowded into a 5 mm drop. In practice, most food and pharmaceutical microbiology labs run spread or pour plates as their primary quantitative method and keep Miles & Misra as the troubleshooting tool.
Drop plate (Herigstad 2001) in detail
The drop plate is the modern successor to Miles & Misra. The Herigstad, Hamilton & Heersink paper in Journal of Microbiological Methods 44(2):121-129 used computer simulation and regression to derive the standard deviation of the drop-plate density estimate as a function of beaker volume, dilution factor, and plated volume. Their contribution was not the technique (Miles & Misra already existed) but the statistical optimisation of how many drops, what volume per drop, and what dilution factor minimise the final coefficient of variation.
How it works
A 10-fold serial dilution series is prepared and 10 microlitre drops are pipetted onto a single agar plate in a grid, typically 5 by 5 (25 drops) or 4 by 5 (20 drops), of one dilution per plate. Drops are allowed to soak in and the plate is incubated to small colonies (16 to 24 hours for fast-growing aerobes). Only drops with 3 to 30 colonies are counted; drops above 30 are excluded for colony merging and drops below 3 for Poisson error. The paper's key finding is that the drop-plate standard-deviation expression is the same as for spread plating, which means the drop plate achieves equivalent precision when its volume and dilution factor are chosen correctly.
When the drop plate wins
The drop plate wins on three criteria: throughput, consumables cost, and incubator space. A high-throughput biofilm or antimicrobial susceptibility screen that would need 400 spread plates (4 dilutions × 100 samples) fits on 100 drop plates (1 dilution × 100 samples, 25 replicates each). The Herigstad 2001 design is the default method in the Montana State University Center for Biofilm Engineering, which is why it appears in every biofilm MIC and MBEC protocol. A 2013 Veterinary Research Forum comparison on Lactobacillus casei and Salmonella Typhimurium reported that the drop method could replace spread plating in routine viable counting without loss of precision. What the drop plate still does not get is pharmacopoeial acceptance, which keeps the spread plate in the release-test slot.
Pros and cons
Miles & Misra
Advantages
- Six to eight dilutions on one plate; always hits the countable window regardless of starting density.
- Lowest agar and plastic consumption of the three methods per final density estimate.
- Eighty-eight years of literature; universally understood by microbiologists and reviewers.
- Reads in 16 to 24 hours because drops are small and colonies stay separated.
Disadvantages
- No pharmacopoeial recognition; results are research-grade or in-house only.
- Precision pipetting of 20 microlitre drops is the dominant error source.
- Each drop counts only 2 to 20 colonies, so single-drop Poisson error is high; needs three replicate drops per dilution.
- Must be read before colonies grow enough to merge across sector boundaries.
Spread plate
Advantages
- Pharmacopoeial reference method (Ph. Eur. 2.6.12, USP 61, FDA BAM Chapter 3).
- Highest colonies per data point (25 to 250), so the lowest Poisson relative error per plate.
- Colonies have full surrounding agar for morphology, picking, and presumptive identification.
- Compatible with every laboratory information management system that logs "plate count".
Disadvantages
- Highest plate count per sample: three to five plates to bracket an unknown dilution.
- Highest agar, plastic, and incubator space consumption of the three.
- Spreading can damage delicate or fast-moving cells (Proteus, swarmers) if the surface is too moist.
- Slower read because colonies grow to a size that fills a 90 mm plate (18 to 48 hours).
Drop plate (Herigstad)
Advantages
- Statistically optimised: Herigstad 2001 derives the dilution and volume that minimise CV.
- Twenty to 25 built-in replicates per dilution on one plate; CV drops to 10 to 15 percent routinely.
- Highest throughput of the three; fits automated multi-channel pipetting and plate-stamping arrays.
- Lowest reagent cost per data point for screening scale (biofilm MIC, antimicrobial screens).
Disadvantages
- Requires an initial density estimate so one dilution can hit the countable window.
- Not pharmacopoeial; QC labs may need a parallel spread plate on release batches.
- Agar surface must be bone dry; a slightly wet plate lets drops merge and the grid collapses.
- Operators new to the layout tend to under-replicate; three-drop minimum must be enforced in SOPs.
Which should you choose?
Three constraints dominate: regulatory scope, whether the density is known within a factor of 10, and how many samples you are counting per day.
Release testing for a regulated product
The number goes on a certificate of analysis or into a regulatory filing. Ph. Eur. 2.6.12, USP <61>, and FDA BAM Chapter 3 all call for spread or pour plating, not drop plating, and reviewers are familiar with the 25 to 250 CFU per plate countable range.
Choose spread plateUnknown starting density, environmental sample
Natural water, soil, wastewater, or a fermentation sample from an unusual timepoint where the density could be 10³ or 10⁹ CFU per mL. Miles & Misra covers that 10⁴-fold range on one plate so at least one sector lands in the countable window on the first attempt.
Choose Miles & MisraHigh-throughput screening or biofilm MIC
100 to 1,000 samples per week in a biofilm MIC, antimicrobial susceptibility screen, or probiotic release count where the density is already known within a factor of 10 from the previous passage.
Choose drop plateTeaching lab or microbiology training
Students need a result on their first attempt without the instructor knowing the exact density of the test culture. The Miles & Misra layout guarantees a readable sector regardless of how well the student prepared the dilution series, which is why it is the default teaching method in UK, Australian, and Indian undergraduate microbiology practicals.
Choose Miles & MisraReal-world use cases
Typical setups where microbiology labs have converged on one method or another.
USP <61> non-sterile product testing
Spread plate on soybean casein digest agar (TSA) at 30 to 35°C, 25 to 250 CFU per plate countable range, three to five plates per sample covering adjacent dilutions. Required method for release of non-sterile pharmaceuticals.
CDC biofilm reactor MBEC
Drop plate on TSA using the Herigstad 2001 20 µL optimisation. 25 drops per coupon dislodge-and-dilution sample, read at 16 hours. Standard assay in the Montana State University Center for Biofilm Engineering biofilm MBEC protocol.
Potable water heterotrophic count
Miles & Misra on R2A agar at 22°C for 72 hours. Eight-sector layout covers 10 to 10⁶ CFU per mL on one plate, which matches the density range of reservoir, mains, and point-of-use water samples.
Lactobacillus / Bifidobacterium release count
Drop plate or spread plate on MRS or MRS-cysteine agar (anaerobe pouches). Drop plate wins on throughput when a probiotic manufacturer runs 50 to 100 batch release counts per week; spread plate wins when the count is on the certificate of analysis for a regulated dietary supplement.
Need to plan the dilution series before you plate?
The cell counting calculator handles the hemocytometer and CFU plate-count maths in one tool. Enter your dilution factor and counted colonies or cells and get density per millilitre without the paper notebook.
Open the Cell Counting CalculatorCost and consumable considerations
The spread plate uses more plates per sample. Miles & Misra and the Herigstad drop plate use fewer. At scale, that moves the per-sample cost by 3 to 5x because an operator-hour, 90 mm petri dish, and 25 mL of agar add up quickly across 50 to 1,000 samples per week. The right comparison is cost per final CFU/mL estimate, not cost per plate.
A single CFU count from a spread plate typically costs 2 to 3 petri dishes (bracketing 3 dilutions), 50 to 75 mL of agar, three sterile L-spreaders or a bag of beads, and 15 to 20 operator minutes. Over 100 samples that is 200 to 300 plates, 5 to 7.5 L of agar, and 25 to 33 operator-hours.
A Miles & Misra count costs one sectored plate, about 25 mL of agar, no spreader, and 15 minutes of operator time. The drop plate (Herigstad 20-drop layout) costs one plate, 25 mL of agar, and 10 to 15 operator minutes. At 100 samples per week the drop plate and Miles & Misra each save 70 to 100 plates and 15 to 25 operator-hours per week versus a straight spread-plate workflow. On a research or probiotic QC budget that is one FTE-week per year reallocated to higher-value work.
| Cost component (per 100 samples) | Miles & Misra | Spread plate | Drop plate |
|---|---|---|---|
| Petri dishes | ~100 | 250 to 300 | ~100 |
| Agar volume | ~2.5 L | ~6 to 7.5 L | ~2.5 L |
| Spreaders / beads | 0 | 300 spreaders or 3 bead bags | 0 |
| Operator time | ~25 h | ~30 h | ~22 h |
| Per-100-sample consumable cost (indicative) | ~$80 to $120 | ~$220 to $340 | ~$80 to $120 |
Equipment and consumables landscape
CFU methods are inherently low-tech. Vendors in this space supply the plates, the dispensers, and the colony counters; the method itself is a protocol decision.
Automated colony counters
- Synbiosis ProtoCOL 3 / ProtoCOL automated colony counter: benchtop imaging colony counter with validated algorithms for spread and drop plates; counts 150 plates per hour, exports to LIMS, widely deployed in pharma QC.
- Interscience Scan 4000 and Scan 1200: image-based colony counters with pharmacopoeial compliance (21 CFR Part 11), fits spread and drop plate workflows from a single platform.
- Bio-Rad Quantity One / ChemiDoc: general-purpose imaging workstations used with open-source colony counting plugins; cheaper than a dedicated counter when imaging capacity is already in the lab.
Plate dispensers and liquid handling
- Hamilton Microlab STAR / STARlet: automated liquid handlers programmed for serial dilution and drop plating; the Montana State University Center for Biofilm Engineering runs Herigstad drop plates on a Hamilton for high-throughput biofilm work.
- INTEGRA Biosciences VIAFLO electronic pipettes: 8- and 16-channel electronic pipettes that dispense repeatable 10 µL drops across a grid; the practical default for manual Herigstad drop plates.
- Eppendorf Research plus and Xplorer plus: single-channel electronic multistep dispensing for 20 µL Miles & Misra drops; eliminates the pipetting-precision error that dominates the single-drop CV.
Agar and plates
- Thermo Fisher Scientific Oxoid: pre-poured TSA, SDA, MacConkey, R2A, and MRS plates sized for 90 mm petri dishes; the industry default for a pharmacopoeial spread plate.
- Merck / Sigma-Aldrich: dehydrated media, pre-poured plates, and the historical supplier for Ph. Eur. media formulations used in USP <61>.
- BD Difco / BBL: the other standard plate and dehydrated-media line for aerobic plate counts, with the published quality certificates FDA BAM Chapter 3 expects to see.
Frequently asked questions
What is the difference between Miles & Misra, spread plate, and drop plate methods?
Which CFU method is most accurate?
Why does Miles & Misra use sectored plates?
What is the countable range for the drop plate method?
When should I use spread plate over Miles & Misra?
Is drop plate as accurate as spread plate?
How many colonies should I count per plate?
What is the difference between pour plate and spread plate?
Do I need technical replicates with the drop plate method?
Resources and references
- Miles, A. A., Misra, S. S., & Irwin, J. O. (1938). The estimation of the bactericidal power of the blood. Journal of Hygiene, 38(6), 732–749. — the original 1938 paper that introduced the sectored surface viable count still known as the Miles & Misra method.
- Herigstad, B., Hamilton, M., & Heersink, J. (2001). How to optimize the drop plate method for enumerating bacteria. Journal of Microbiological Methods, 44(2), 121–129. — the statistical optimisation of the drop plate method; derives the volume and dilution factor that minimise variance and shows the same expression applies to spread plating.
- Hedges, A. J. (2002). Estimating the precision of serial dilutions and viable bacterial counts. International Journal of Food Microbiology, 76(3), 207–214. — error propagation through serial dilutions for both standard and Miles & Misra plate counts; shows that replicating the final plating step beats averaging across dilutions.
- FDA Bacteriological Analytical Manual (BAM) Chapter 3: Aerobic Plate Count (2026 edition). — the pharmacopoeial reference method for spread and pour plate counts; the source of the 25 to 250 CFU per plate countable range cited throughout this comparison.