Assessment and certificate
Interpret trends, calculate process quantities and put calibration steps in order.
You will be able to
- Select independent evidence before acting on a change.
- Check the time, units and flow basis of a calculation.
- Interpret agreement and disagreement between instruments.
Review RQ, t90 and kLa before the core assessment.
Student assessment
Answer 15 questions drawn from a bank of 30. Each question carries one mark. At least 11 correct answers are needed to reach 70%. Your score and explanations appear when you submit.

The core sensors from this course on one benchtop bioreactor. Before you start the assessment, name what each one measures and how it can mislead.
- pH electrode (black connector cap)
- Optical dissolved oxygen probe
- Pt100 in its thermowell
- Headspace pressure sensor
- Feed balance
Quick check · 2 questions
Professional case assessment
Each attempt draws 10 cases from a bank of 20. A score of 7 out of 10 passes this part. Both passes are required for a Professional certificate.
Quick check · 2 questions
Print your certificate
Pass the core assessment, then enter the name to appear on your certificate. Your name stays on this page. Print from this session before closing the page.
Sources and further reading
- J. H. T. Luong et al., “Instrumentation and Analytical Methods”, Comprehensive Biotechnology, 3rd ed., edited by M. Moo-Young, Elsevier, 2019, ch. 2.60.
- W. Klöckner and J. Büchs, “Shake-Flask Bioreactors”, Comprehensive Biotechnology, 3rd ed., edited by M. Moo-Young, Elsevier, 2019, ch. 2.24.
- C. Herwig, R. Pörtner and J. Möller (eds.), Digital Twins: Applications to the Design and Optimization of Bioprocesses, Springer, 2021, chs. 3 and 4 on control strategies and state estimation.
BioProcess Tools
Awarded to
for completing
Sensors and Process Monitoring for Bioreactors
Review and apply
Revisit the opening decision
Module flashcards
Professional review cards
Cross-module review
Confidence by objective
Choose low to bring the related cards forward for review.
BioProcess Tools · Module summary
Assessment and certificate
Key relationships
- RQ = molar CER / molar OUR.
- First-order t90 = ln(10) × τ.
- Net μ = ln[(X₂V₂)/(X₁V₁)]/Δt when losses are negligible.
Faults and checks
| Observation | Interpretation or check |
|---|---|
| Offline and online values disagree | Check sample timestamps, units and measurands. |
| Plausible value on two displays | Verify independent calibration before trusting agreement. |
Numbers to remember
- Student assessment contains 15 questions. A pass requires 11 correct answers.
- Module practice and confidence ratings support review. They do not determine the quiz result.