Aseptic operations
Set up a cabinet, design a representative sample and follow a closed transfer from the source to the receiving vessel.
You will be able to
- Protect the cabinet work zone during a transfer.
- Collect a representative, traceable sample and account for the liquid withdrawn.
- Choose a transfer and connection method compatible with the equipment and procedure.
Read the working zone
Before starting, check cabinet certification, sash height and airflow status. Class II designs can have nominal inward velocities around 0.4–0.5 m/s depending on type. Keep both grilles clear and place open work at least about 10 cm inside the front grille. A horizontal clean bench protects the work and has a different airflow pattern from a personnel-containment cabinet.

Room air is drawn in at the front grille while HEPA-filtered air falls over the work surface as a downflow curtain. Keep both grilles clear and work well back from the front.
- Downflow HEPA filter (section)
- HEPA-filtered downflow splitting to the front and rear grilles
- Room air drawn in under the sash
- Front intake grille
- Rear grille
- Work placed well back from the front grille
Perform hand hygiene and use the required gown and gloves. On appropriate clean surfaces, 70% alcohol helps denature proteins and acts on many vegetative organisms. Keep the surface wet for the validated contact time. Alcohol does not kill bacterial spores. Use the approved sporicidal agent where the procedure requires it.
Quick check · 2 questions
Take a representative sample
Calculate the standing volume in both the tube and the valve. In a flow-through sample line, 3–5 line volumes is a common starting rule for a qualified purge. A measured 2.38 mL line at a three-volume purge sends 7.14 mL to the waste receiver. Confirm the site’s purge volume and port method. Add the analytical sample separately when recording the overall volume removed.

Culture standing in the side-port sampler and its line is not representative of the vessel. Purge it to the waste bottle first, then take the analytical sample.
- Sample valve on the vessel port
- Dead-leg volume in the valve and line
- Pinch clamp closed on the sample branch
- Sample bottle
- Purge waste bottle
Quick check · 2 questions
Keep transfers closed
Verify the identity and condition of both ends of the transfer. Match weldable tubing to the qualified welder, or check full engagement of the specified aseptic connector. Keep open flames outside a biosafety cabinet. For suitable pressure-rated bench or pilot vessels, 0.2–0.5 bar gauge is an illustrative headspace overpressure. Check the actual setpoint, relief path and pressure trend during cooling and sampling. Flexible single-use bags may have much lower limits.

Medium moves from the bag to the vessel through one closed tubing path: a sterile connection, a peristaltic pump and a protected headplate port. Check both ends before you start.
- Media bag
- Mated sterile connector
- Peristaltic pump head
- Inoculation septum on the headplate
Quick check · 2 questions
Practice and summary
Revisit your opening decision
BSC · first air · dead volume · closed transfer
Module flashcards
Confidence by objective
A low rating brings related cards forward for review.
Sources and further reading
Books
- W. L. Hochfeld. Producing Biomolecular Substances with Fermenters, Bioreactors, and Biomolecular Synthesizers. CRC Press, 2006. Chapters 8 and 9, fermenters and bioreactors.
- J. H. T. Luong, K. A. Mahmoud and K. B. Male. “Instrumentation and Analytical Methods.” In M. Moo-Young (ed.), Comprehensive Biotechnology, 3rd ed. Elsevier, 2019. Chapter 2.60.
- R. Pörtner (ed.). Biopharmaceutical Manufacturing: Progress, Trends and Challenges. Springer Nature Switzerland, 2023. Chapter 15, advanced therapies.
Guidance and background
- US CDC. Guideline for Disinfection and Sterilization in Healthcare Facilities, chemical disinfectants and alcohols.
- CDC and NIH. Biosafety in Microbiological and Biomedical Laboratories, 6th ed. Appendix A, biological safety cabinets.
Related BioProcess Tools pages
BioProcess Tools · Aseptic technique · Module 3
Aseptic operations
Takeaways
- Confirm cabinet status and working sash height. Keep grilles clear and the route of first air to exposed critical sites unobstructed.
- Use the locally approved gowning, hand hygiene and disinfection procedures. Wet contact time is specific to the product and method.
- Identify the sample and receiver. The collection record should include withdrawal time, the approved purge and the collected volume.
- For an addition, verify the material and choose a qualified compatible connection. Record the intervention and restore the boundary.
Check your understanding
- Protect the cabinet work zone during a transfer.
- Collect a representative, traceable sample and account for the liquid withdrawn.
- Choose a transfer and connection method compatible with the equipment and procedure.