Course overviewCourse progress
Module 02 of 06

Sterilisation

Compare the exposure at the load with the chamber record, then interpret the evidence for a membrane or a single-use assembly.

You will be able to

  • Relate load temperature, exposure time and survivor probability.
  • Interpret the different evidence supplied by a cold-point trace and by indicators.
  • Choose appropriate integrity evidence for membranes and single-use assemblies.
One-minute decision

Read the thermal exposure

Use the cold-point trace. One D-value is the time for a tenfold decrease at a given temperature. Raising temperature by one z-value cuts D tenfold. Conventional F₀ adds equivalent minutes at 121.1 °C using z = 10 °C. A sterility assurance level (SAL) of 10⁻⁶ means that the estimated probability of an individual item remaining non-sterile after the validated process is at most one in a million.

A front-loading laboratory steam steriliser with its round door open and the right side cut away. A wire basket holds media bottles with loosened caps under foil, a pack in blue steriliser wrap with indicator tape and a biological indicator vial standing beside the largest bottle. An orange thermocouple lead runs from a gland in the chamber back wall into the largest bottle, where the probe tip sits in the lower centre of the liquid. A pressure gauge reading in kPa absolute and a control display sit above the door.
3D figureA loaded steam steriliser

The cold-point trace comes from a thermocouple in the slowest-heating item, here the largest liquid bottle. Caps on liquid bottles are loosened so that air and pressure can escape during the cycle.

  1. Load thermocouple, tip in the slowest-heating bottle
  2. Biological indicator at the mapped challenge location
  3. Pack in steriliser wrap with indicator tape
  4. Media bottle with loosened cap and foil
  5. Chamber pressure gauge (kPa absolute)
  6. Cycle display and controls

For an illustrative burden of 10⁶ per item, twelve decimal reductions take the expected count to 10⁻⁶. The familiar overkill example of F₀ = 12 min assumes D₁₂₁.₁ = 1 min. A 1.5 min D-value needs 18 equivalent minutes for the same twelve reductions. In continuous high-temperature, short-time (HTST) processing, heated medium passes through a holding tube and is cooled promptly. In the medium example below, a Del requirement of 34.5 and an illustrative death-rate constant of 56.5 min⁻¹ at 140 °C give about 37 s in an ideal holding tube. Verify the shortest residence time and coldest material as well as the nominal temperature.

Quick check · 2 questions

Compare chamber and load

Choose the validated cycle for the load. Gravity displacement moves air out through the drain. Pulsed pre-vacuum removes air before steam exposure and helps steam penetrate porous loads. A Bowie-Dick test checks air removal and steam penetration in a dynamic-air-removal steriliser before use. A passing pattern cannot release a liquid load.

Inside a cut-away steriliser chamber, a 2 L bottle with a loosened cap and a wrapped pack sit on a wire basket, both shown in section. The liquid in the bottle shades from warm orange at the walls to cool blue at its lower centre, where the tip of a probe sits. The probe lead runs through the air to a gland in the chamber wall. The pack contains layers of folded material with a pale blue air pocket in the middle. Faint orange arrows of steam point at both items, and a sensor sits in the drain line below the chamber.
3D figureChamber temperature versus load temperature

Steam condenses on the outside of the load first, so the drain sensor can read the set temperature while the centre of a bottle or pack is still cooler. Trapped air slows heating further.

  1. Condensing steam heating the load surfaces
  2. Trapped air pocket inside a wrapped porous pack
  3. Load probe, tip at the bottle's cold point
  4. Chamber temperature sensor in the drain line

Use vented or loosened caps on liquid bottles compatible with that cycle. A typical fill limit is about two-thirds to three-quarters, subject to the approved load specification. Leave space for steam circulation, allow slow exhaust for liquids and monitor the slowest-heating bottle. In a steel vessel, clean before SIP and check that low points drain condensate.

At the mapped challenge location, a steam biological indicator (BI) may contain Geobacillus stearothermophilus spores. A typical D₁₂₁ is 1.5–3 min. Use the lot certificate and the indicator instructions, commonly 55–60 °C incubation, with an unexposed positive control. Interpret the result with the load temperature record.

Quick check · 2 questions

Interpret a filter integrity test

Read the bacterial-retention evidence for the membrane. ASTM F838 uses a Brevundimonas diminuta challenge of 10⁷ CFU/cm² of effective membrane area. A routine integrity reading is compared with a product- and wetting-fluid-specific limit correlated with retention. A 0.2 µm rating alone does not establish removal of mycoplasma or dissolved endotoxin.

A stainless filter housing on a bench stand with a quarter of the bowl cut away to show the wetted, pleated membrane cartridge inside. A pressure gauge sits on the housing head. Tubing runs from a gas regulator to a benchtop integrity test instrument with a pressure trace on its screen, from the instrument to the housing inlet, and from the housing outlet into an open beaker.
3D figureFilter integrity test set-up

The test instrument pressurises the upstream side of the wetted cartridge and records the reading, which is compared with a limit for that product and wetting fluid.

  1. Pleated sterilising-grade membrane (cutaway)
  2. Stainless filter housing
  3. Upstream pressure gauge
  4. Inlet gas regulator
  5. Integrity test instrument

For sterile medicinal-product filtration, Annex 1 §8.87 calls for a pre-use post-sterilisation integrity test (PUPSIT) and a post-use integrity test. Where PUPSIT is impracticable, Annex 1 provides for a documented risk assessment and appropriate alternative controls. Read the approved configuration and the test record together.

Quick check · 2 questions

Accept a single-use assembly

Match the single-use assembly and irradiation lot to the supplier certificate. A documented gamma-dose range might be 25–50 kGy. Confirm the qualified minimum and the product’s tolerated maximum for that assembly. Inspect both layers of packaging and the connector. A tear in the sterile barrier puts the assembly on hold even when the irradiation certificate is satisfactory.

A gamma-irradiated single-use assembly on a blue-grey tray: a flat pillow bag with three welded ports, silicone tubing to a capped sterile connector seated on its line, a capsule vent filter and a capped sampling line, with blue pinch clamps. Beside the tray lie the opened outer and inner pouches, peeled back at their chevron seals, with a label carrying a barcode and a red irradiation indicator dot.
3D figureAccepting a single-use assembly

Match the irradiation indicator and lot to the certificate, then inspect both layers of packaging and every connector before use.

  1. 2D pillow bag with welded ports
  2. Sterile connector, capped
  3. Pinch clamp on the transfer line
  4. Vent filter
  5. Irradiation indicator on the pack label
  6. Opened outer and inner pouches
Quick check · 2 questions

Practice and summary

Revisit your opening decision

D-value · z-value · F₀ · Del factor · SIP · bubble point

Module flashcards
Confidence by objective

A low rating brings related cards forward for review.

Sources and further reading

Books

  • P. Verma (ed.). Industrial Microbiology and Biotechnology: A New Horizon of the Microbial World. Springer Nature Singapore, 2024. Chapters 4 and 7, process development and thermal death kinetics.
  • W. L. Hochfeld. Producing Biomolecular Substances with Fermenters, Bioreactors, and Biomolecular Synthesizers. CRC Press, 2006. Chapters 8 and 9, fermenters and bioreactors.
  • C.-F. Mandenius (ed.). Bioreactors: Design, Operation and Novel Applications. Wiley-VCH, 2016. Chapters 1 and 9, vessel design and single-use systems.

Guidance and background

Related BioProcess Tools pages

BioProcess Tools · Aseptic technique · Module 2

Sterilisation

Takeaways

Check your understanding