Sterilisation
Compare the exposure at the load with the chamber record, then interpret the evidence for a membrane or a single-use assembly.
You will be able to
- Relate load temperature, exposure time and survivor probability.
- Interpret the different evidence supplied by a cold-point trace and by indicators.
- Choose appropriate integrity evidence for membranes and single-use assemblies.
Read the thermal exposure
Use the cold-point trace. One D-value is the time for a tenfold decrease at a given temperature. Raising temperature by one z-value cuts D tenfold. Conventional F₀ adds equivalent minutes at 121.1 °C using z = 10 °C. A sterility assurance level (SAL) of 10⁻⁶ means that the estimated probability of an individual item remaining non-sterile after the validated process is at most one in a million.

The cold-point trace comes from a thermocouple in the slowest-heating item, here the largest liquid bottle. Caps on liquid bottles are loosened so that air and pressure can escape during the cycle.
- Load thermocouple, tip in the slowest-heating bottle
- Biological indicator at the mapped challenge location
- Pack in steriliser wrap with indicator tape
- Media bottle with loosened cap and foil
- Chamber pressure gauge (kPa absolute)
- Cycle display and controls
For an illustrative burden of 10⁶ per item, twelve decimal reductions take the expected count to 10⁻⁶. The familiar overkill example of F₀ = 12 min assumes D₁₂₁.₁ = 1 min. A 1.5 min D-value needs 18 equivalent minutes for the same twelve reductions. In continuous high-temperature, short-time (HTST) processing, heated medium passes through a holding tube and is cooled promptly. In the medium example below, a Del requirement of 34.5 and an illustrative death-rate constant of 56.5 min⁻¹ at 140 °C give about 37 s in an ideal holding tube. Verify the shortest residence time and coldest material as well as the nominal temperature.
Quick check · 2 questions
Compare chamber and load
Choose the validated cycle for the load. Gravity displacement moves air out through the drain. Pulsed pre-vacuum removes air before steam exposure and helps steam penetrate porous loads. A Bowie-Dick test checks air removal and steam penetration in a dynamic-air-removal steriliser before use. A passing pattern cannot release a liquid load.

Steam condenses on the outside of the load first, so the drain sensor can read the set temperature while the centre of a bottle or pack is still cooler. Trapped air slows heating further.
- Condensing steam heating the load surfaces
- Trapped air pocket inside a wrapped porous pack
- Load probe, tip at the bottle's cold point
- Chamber temperature sensor in the drain line
Use vented or loosened caps on liquid bottles compatible with that cycle. A typical fill limit is about two-thirds to three-quarters, subject to the approved load specification. Leave space for steam circulation, allow slow exhaust for liquids and monitor the slowest-heating bottle. In a steel vessel, clean before SIP and check that low points drain condensate.
At the mapped challenge location, a steam biological indicator (BI) may contain Geobacillus stearothermophilus spores. A typical D₁₂₁ is 1.5–3 min. Use the lot certificate and the indicator instructions, commonly 55–60 °C incubation, with an unexposed positive control. Interpret the result with the load temperature record.
Quick check · 2 questions
Interpret a filter integrity test
Read the bacterial-retention evidence for the membrane. ASTM F838 uses a Brevundimonas diminuta challenge of 10⁷ CFU/cm² of effective membrane area. A routine integrity reading is compared with a product- and wetting-fluid-specific limit correlated with retention. A 0.2 µm rating alone does not establish removal of mycoplasma or dissolved endotoxin.

The test instrument pressurises the upstream side of the wetted cartridge and records the reading, which is compared with a limit for that product and wetting fluid.
- Pleated sterilising-grade membrane (cutaway)
- Stainless filter housing
- Upstream pressure gauge
- Inlet gas regulator
- Integrity test instrument
For sterile medicinal-product filtration, Annex 1 §8.87 calls for a pre-use post-sterilisation integrity test (PUPSIT) and a post-use integrity test. Where PUPSIT is impracticable, Annex 1 provides for a documented risk assessment and appropriate alternative controls. Read the approved configuration and the test record together.
Quick check · 2 questions
Accept a single-use assembly
Match the single-use assembly and irradiation lot to the supplier certificate. A documented gamma-dose range might be 25–50 kGy. Confirm the qualified minimum and the product’s tolerated maximum for that assembly. Inspect both layers of packaging and the connector. A tear in the sterile barrier puts the assembly on hold even when the irradiation certificate is satisfactory.

Match the irradiation indicator and lot to the certificate, then inspect both layers of packaging and every connector before use.
- 2D pillow bag with welded ports
- Sterile connector, capped
- Pinch clamp on the transfer line
- Vent filter
- Irradiation indicator on the pack label
- Opened outer and inner pouches
Quick check · 2 questions
Practice and summary
Revisit your opening decision
D-value · z-value · F₀ · Del factor · SIP · bubble point
Module flashcards
Confidence by objective
A low rating brings related cards forward for review.
Sources and further reading
Books
- P. Verma (ed.). Industrial Microbiology and Biotechnology: A New Horizon of the Microbial World. Springer Nature Singapore, 2024. Chapters 4 and 7, process development and thermal death kinetics.
- W. L. Hochfeld. Producing Biomolecular Substances with Fermenters, Bioreactors, and Biomolecular Synthesizers. CRC Press, 2006. Chapters 8 and 9, fermenters and bioreactors.
- C.-F. Mandenius (ed.). Bioreactors: Design, Operation and Novel Applications. Wiley-VCH, 2016. Chapters 1 and 9, vessel design and single-use systems.
Guidance and background
- ASTM International. ASTM F838-20, Standard Test Method for Determining Bacterial Retention of Membrane Filters Utilized for Liquid Filtration.
- European Commission. EU GMP Annex 1 (2022), §8.87, integrity testing of sterilising filters.
- US CDC. Guideline for Disinfection and Sterilization in Healthcare Facilities, sterilizing practices.
- US FDA. Sterile Drug Products Produced by Aseptic Processing (2004). Guidance on filtration and process validation.
- ISO 11138-3:2017. Biological indicators for moist heat sterilization processes.
- ISO 11140-1:2014. Chemical indicators, general requirements.
- ISO 11137-1. Radiation sterilisation, process development, validation and routine control.
Related BioProcess Tools pages
BioProcess Tools · Aseptic technique · Module 2
Sterilisation
Takeaways
- D is the time for one decimal reduction at a stated temperature. z is the temperature change for a tenfold change in D.
- F₀ integrates local temperature over time using 121.1 °C and z = 10 °C. Del uses the natural logarithm of the ratio of initial to expected survivors.
- Review the validated load configuration, mapped cold location, indicators and cycle records together. Steam traps discharge condensate.
- Use the specified membrane, wetting fluid and integrity criterion. Verify single-use assembly identity, release, packaging and expiry.
Check your understanding
- Relate load temperature, exposure time and survivor probability.
- Interpret the different evidence supplied by a cold-point trace and by indicators.
- Choose appropriate integrity evidence for membranes and single-use assemblies.