Temperature, pH and dissolved oxygen
Calibrate two probes and diagnose six instrument faults from their signal traces.
You will be able to
- Interpret resistance and temperature lag.
- Relate pH and oxygen to electrical or optical signals.
- Distinguish sensor faults using independent evidence.
Temperature at the sensing point
A Pt100 is a platinum resistance thermometer with a nominal resistance of 100 Ω at 0 °C. Lead resistance matters, especially with two-wire wiring. A thermowell isolates the element from the broth. Poor seating or a thick well adds response lag.

Four 12 mm probes for standard headplate ports, each with a different sensing tip and connector head. The Pt100 sits inside a thermowell, which isolates the element from the broth.
- Pt100 element at the tip of the thermowell
- pH glass membrane bulb
- Reference junction
- Optical DO sensor cap
- Polarographic DO membrane cap
- Connector head on the threaded fitting
The temperature loop compares the measurement with its setpoint and changes heat exchange through the jacket. The Q10 ratio describes the change in a biological rate over 10 °C. A Q10 of 2 implies about a 7% rise for 1 °C within that local model. Near a growth limit, the same extrapolation can fail.
Quick check · 2 questions
Two-buffer pH calibration
A hydrated glass membrane develops a potential that depends on hydrogen ion activity. A reference half-cell completes the circuit through a liquid junction. In the polarity used here the ideal slope is −2.303 RT/F, about −59.16 mV per pH unit at 25 °C. The intercept at pH 7 is reported as the offset.

Confirm the neutral buffer first, then rinse and blot the electrode before the second buffer. Read each buffer at its actual temperature.
- Glass bulb immersed in the first buffer
- pH 7 buffer, confirmed first
- pH 4 buffer for the second point
- Meter reading with temperature compensation
- Wash bottle for rinsing between buffers
Use fresh buffers spanning the operating region. Read each buffer’s pH at its actual temperature. For the meter in this exercise, confirm the neutral buffer first, rinse and blot gently, then confirm the second buffer after stabilisation. Temperature compensation adjusts electrode sensitivity. The sample itself can change pH with temperature.
Choose the sensor for the run
A blocked junction can cause erratic offsets. Ageing can reduce slope. A cracked bulb after autoclaving can give a plausible but unresponsive signal. An ISFET senses pH with a field-effect transistor and still needs a stable reference. Optical spots use a pH-sensitive dye read through a window. The dye response flattens near its limits. One single-use example covers roughly pH 6.5 to 8.5. Check the validated range of each design.
Quick check · 2 questions
Two ways to sense dissolved oxygen

Left, an optical probe reads the luminescence of a dye spot through a window. Right, a Clark probe reduces oxygen that has diffused through its membrane at a biased cathode.
- Luminophore sensor spot facing the broth
- Optical window
- Light guide to the LED and detector
- Gas-permeable membrane
- Cathode behind the membrane
- Silver anode in the electrolyte
| Principle | Readout and upkeep |
|---|---|
| Clark polarographic | Oxygen diffuses through a membrane and is reduced at a biased cathode. The current depends on diffusion. Polarise before calibration. Check electrolyte, membrane and local flow. Fouling slows response and ageing can shift the baseline. |
| Optical quenching | Oxygen shortens the excited-state lifetime of a dye. A phase shift under modulated illumination gives tan(φ) = ωτ. Lifetime readout reduces sensitivity to intensity changes. Photobleaching, deposits and interfering emission still matter. |
For this idealised Stern–Volmer model, τ₀ is the lifetime at zero oxygen. Real foils may require a multi-site calibration. At 100% air saturation the broth is equilibrated with air at the stated temperature and pressure. A percentage of air saturation is distinct from the mole percentage of oxygen in the gas.
After sterilisation
For the Module 2 procedure, the Clark probe starts unpolarised. After steam sterilisation, cool the installed system to process temperature and pressure. Complete the specified polarisation period. Confirm a stable oxygen-free zero, then a stable air-saturated span at the intended mixing conditions. Other procedures may keep the probe polarised during sterilisation or assign the two points in a different order. Follow the procedure for the installed system. Keep sterile boundaries intact. Verify the span independently. Use the specified post-sterilisation check for a factory-calibrated disposable spot.
Quick check · 2 questions
Sensor doctor
Compare the process trace with the independent reference or the recorded calibration event. Choose both the likely fault and the corrective action. Six cases contribute to your score.
Quick check · 2 questions
Single-use qualification and interference
Gamma irradiation can change the response of an optical indicator. A calibration measured before irradiation may shift after the validated dose. Link the lot and irradiation record to the installed sensor. Compare a post-treatment reference point and its uncertainty with the permitted decision error.
When the culture fluoresces
A fluorescent product can overlap a dye’s excitation or emission band. In a reported yellow fluorescent protein example the optical DO signal reached about 300% air saturation. Check gas composition, pressure and a different sensing principle. Run matrix controls with and without the fluorescent product before using the signal for a cascade.
Qualify the installed system
General practice separates installation qualification (IQ), operational qualification (OQ) and performance qualification (PQ). IQ checks the correct probe, port, serial number and software channel. OQ challenges the span, alarms and failure handling. PQ tests repeatable performance in representative broth across intended runs.
Record reference identity, traceability, temperature, acceptance limits, as-found and as-left results, operator, time and maintenance. An out-of-tolerance result requires review of affected data since the last acceptable check.
Quick check · 2 questions
Sources and further reading
- J. H. T. Luong et al., “Instrumentation and Analytical Methods”, Comprehensive Biotechnology, 3rd ed., edited by M. Moo-Young, Elsevier, 2019, ch. 2.60.
- R. Pörtner (ed.), Biopharmaceutical Manufacturing: Progress, Trends and Challenges, Springer, 2024 print edition, chs. 1 and 2 on single-use systems and monitoring.
- C.-F. Mandenius (ed.), Bioreactors: Design, Operation and Novel Applications, Wiley-VCH, 2016, section 2.6 (optical sensing and fluorescent-product interference).
- B. Hitzmann (ed.), Bioprocess Monitoring and Control, MDPI, 2020, Noll et al., “Modeling and Exploiting Microbial Temperature Response”.
Review and apply
Revisit the opening decision
Module flashcards
Professional review cards
Confidence by objective
Choose low to bring the related cards forward for review.
BioProcess Tools · Module summary
Temperature, pH and dissolved oxygen
Key relationships
- Near 0 °C, ΔT ≈ ΔR / (0.385 Ω/°C).
- Ideal pH slope = −2.303 RT/F. Optical phase gives tan(φ) = ωτ.
Faults and checks
| Observation | Interpretation or check |
|---|---|
| Sluggish temperature step | Check well thickness and element seating. |
| Shallow electrode slope | Inspect two stabilised buffers and glass condition. |
| Unexpected DO | Check an independent principle and actual gas conditions. |
Numbers to remember
- Pt100 = 100 Ω at 0 °C.
- Ideal pH slope near 25 °C is −59.16 mV/pH in the course polarity.