Temperature, pH and dissolved oxygen

Calibrate two probes and diagnose six instrument faults from their signal traces.

You will be able to

  • Interpret resistance and temperature lag.
  • Relate pH and oxygen to electrical or optical signals.
  • Distinguish sensor faults using independent evidence.
One-minute decision

Temperature at the sensing point

A Pt100 is a platinum resistance thermometer with a nominal resistance of 100 Ω at 0 °C. Lead resistance matters, especially with two-wire wiring. A thermowell isolates the element from the broth. Poor seating or a thick well adds response lag.

Four 12 mm probes hanging at a slight forward lean from a slanted bar on a dark display board: a Pt100 insert in a thermowell with a window cut open at the tip, a glass pH electrode with its bulb and side junction, an optical dissolved oxygen probe with an orange sensor spot on its cap, and a polarographic dissolved oxygen probe with a membrane cap. Each has its own connector head.
3D figureCore probes at true relative size

Four 12 mm probes for standard headplate ports, each with a different sensing tip and connector head. The Pt100 sits inside a thermowell, which isolates the element from the broth.

  1. Pt100 element at the tip of the thermowell
  2. pH glass membrane bulb
  3. Reference junction
  4. Optical DO sensor cap
  5. Polarographic DO membrane cap
  6. Connector head on the threaded fitting

The temperature loop compares the measurement with its setpoint and changes heat exchange through the jacket. The Q10 ratio describes the change in a biological rate over 10 °C. A Q10 of 2 implies about a 7% rise for 1 °C within that local model. Near a growth limit, the same extrapolation can fail.

Quick check · 2 questions

Two-buffer pH calibration

A hydrated glass membrane develops a potential that depends on hydrogen ion activity. A reference half-cell completes the circuit through a liquid junction. In the polarity used here the ideal slope is −2.303 RT/F, about −59.16 mV per pH unit at 25 °C. The intercept at pH 7 is reported as the offset.

A pH electrode held in a stand clamp with its glass bulb in a beaker of pH 7.00 buffer. A second beaker is labelled pH 4.01. A benchtop meter shows 7.00 pH at 25.0 °C, with a wash bottle of deionised water and a folded tissue for blotting beside the buffers.
3D figureTwo-buffer pH calibration

Confirm the neutral buffer first, then rinse and blot the electrode before the second buffer. Read each buffer at its actual temperature.

  1. Glass bulb immersed in the first buffer
  2. pH 7 buffer, confirmed first
  3. pH 4 buffer for the second point
  4. Meter reading with temperature compensation
  5. Wash bottle for rinsing between buffers

Use fresh buffers spanning the operating region. Read each buffer’s pH at its actual temperature. For the meter in this exercise, confirm the neutral buffer first, rinse and blot gently, then confirm the second buffer after stabilisation. Temperature compensation adjusts electrode sensitivity. The sample itself can change pH with temperature.

Choose the sensor for the run

A blocked junction can cause erratic offsets. Ageing can reduce slope. A cracked bulb after autoclaving can give a plausible but unresponsive signal. An ISFET senses pH with a field-effect transistor and still needs a stable reference. Optical spots use a pH-sensitive dye read through a window. The dye response flattens near its limits. One single-use example covers roughly pH 6.5 to 8.5. Check the validated range of each design.

Quick check · 2 questions

Two ways to sense dissolved oxygen

Two probe tips cut in half and dipping into broth that contains oxygen molecules. The optical tip shows a glowing orange luminophore layer behind a clear window, teal excitation light running down a light guide and red luminescence returning to the detector. The Clark tip shows a thin membrane over a gold cathode in a glass body with a thin electrolyte film behind it, a silver anode sleeve with dark chloride bands, and blue electrolyte.
3D figureOptical and Clark DO tips, cut away

Left, an optical probe reads the luminescence of a dye spot through a window. Right, a Clark probe reduces oxygen that has diffused through its membrane at a biased cathode.

  1. Luminophore sensor spot facing the broth
  2. Optical window
  3. Light guide to the LED and detector
  4. Gas-permeable membrane
  5. Cathode behind the membrane
  6. Silver anode in the electrolyte
Principle Readout and upkeep
Clark polarographic Oxygen diffuses through a membrane and is reduced at a biased cathode. The current depends on diffusion. Polarise before calibration. Check electrolyte, membrane and local flow. Fouling slows response and ageing can shift the baseline.
Optical quenching Oxygen shortens the excited-state lifetime of a dye. A phase shift under modulated illumination gives tan(φ) = ωτ. Lifetime readout reduces sensitivity to intensity changes. Photobleaching, deposits and interfering emission still matter.
τ₀ / τ = 1 + K · pO₂

For this idealised Stern–Volmer model, τ₀ is the lifetime at zero oxygen. Real foils may require a multi-site calibration. At 100% air saturation the broth is equilibrated with air at the stated temperature and pressure. A percentage of air saturation is distinct from the mole percentage of oxygen in the gas.

After sterilisation

For the Module 2 procedure, the Clark probe starts unpolarised. After steam sterilisation, cool the installed system to process temperature and pressure. Complete the specified polarisation period. Confirm a stable oxygen-free zero, then a stable air-saturated span at the intended mixing conditions. Other procedures may keep the probe polarised during sterilisation or assign the two points in a different order. Follow the procedure for the installed system. Keep sterile boundaries intact. Verify the span independently. Use the specified post-sterilisation check for a factory-calibrated disposable spot.

Quick check · 2 questions

Sensor doctor

Compare the process trace with the independent reference or the recorded calibration event. Choose both the likely fault and the corrective action. Six cases contribute to your score.

Quick check · 2 questions
Professional

Single-use qualification and interference

Gamma irradiation can change the response of an optical indicator. A calibration measured before irradiation may shift after the validated dose. Link the lot and irradiation record to the installed sensor. Compare a post-treatment reference point and its uncertainty with the permitted decision error.

When the culture fluoresces

A fluorescent product can overlap a dye’s excitation or emission band. In a reported yellow fluorescent protein example the optical DO signal reached about 300% air saturation. Check gas composition, pressure and a different sensing principle. Run matrix controls with and without the fluorescent product before using the signal for a cascade.

Qualify the installed system

General practice separates installation qualification (IQ), operational qualification (OQ) and performance qualification (PQ). IQ checks the correct probe, port, serial number and software channel. OQ challenges the span, alarms and failure handling. PQ tests repeatable performance in representative broth across intended runs.

Record reference identity, traceability, temperature, acceptance limits, as-found and as-left results, operator, time and maintenance. An out-of-tolerance result requires review of affected data since the last acceptable check.

Quick check · 2 questions

Sources and further reading

  • J. H. T. Luong et al., “Instrumentation and Analytical Methods”, Comprehensive Biotechnology, 3rd ed., edited by M. Moo-Young, Elsevier, 2019, ch. 2.60.
  • R. Pörtner (ed.), Biopharmaceutical Manufacturing: Progress, Trends and Challenges, Springer, 2024 print edition, chs. 1 and 2 on single-use systems and monitoring.
  • C.-F. Mandenius (ed.), Bioreactors: Design, Operation and Novel Applications, Wiley-VCH, 2016, section 2.6 (optical sensing and fluorescent-product interference).
  • B. Hitzmann (ed.), Bioprocess Monitoring and Control, MDPI, 2020, Noll et al., “Modeling and Exploiting Microbial Temperature Response”.

Review and apply

Revisit the opening decision
Module flashcards

Professional review cards

Confidence by objective

Choose low to bring the related cards forward for review.

BioProcess Tools · Module summary

Temperature, pH and dissolved oxygen

Key relationships

Faults and checks

ObservationInterpretation or check
Sluggish temperature stepCheck well thickness and element seating.
Shallow electrode slopeInspect two stabilised buffers and glass condition.
Unexpected DOCheck an independent principle and actual gas conditions.

Numbers to remember