Small-scale and noninvasive monitoring
Read respiration through a flask headspace and explore how fill volume changes an oxygen-limited screen.
You will be able to
- Identify when respiration is transfer-limited.
- Select a valid closed-phase gas interval.
- Compare optical coverage with scale-down similarity.
Shake-flask monitor
Compare the oxygen demand of an illustrative E. coli batch with the oxygen transfer capacity of the flask. Look for a plateau while DO is low, a break at glucose exhaustion and a second respiratory peak as acetate is consumed. Traces are educational estimates.

Sensor spots inside the flask bottom are read through the glass by a reader under the shaker tray. A spot reads the culture only while liquid covers it.
- Baffle in the flask base
- Sensor spot on the inside of the glass
- Reader under the shaker tray
- Culture climbing the wall as the flask orbits
- Orbital shaker drive
Quick check · 2 questions
Read respiration in a closed measuring phase
In a closed-phase respiration monitor, a rinse phase restores the inlet gas composition. During a short closed measuring phase, the drop in oxygen partial pressure estimates gas-to-liquid oxygen transfer. A total-pressure change plus the oxygen change can resolve CO₂ under a dry, constant-temperature inert balance. Reopen for the next cycle.

During a short closed phase, the sensor in the cap follows the fall in oxygen partial pressure in the headspace. The valves then open to rinse the headspace with inlet gas. Nitrogen is not shown and molecule counts are illustrative.
- Headspace gas
- Oxygen molecule
- Carbon dioxide molecule
- Gas sensor face in the cap
- Inlet and outlet valves for the rinse phase
- Culture liquid
Use absolute temperature. Pair kPa and litres with R = 8.314 kPa·L/(mol·K). At constant gas volume, temperature and liquid volume, RQ is the CO₂ partial-pressure rise divided by the O₂ partial-pressure fall if gas exchange represents metabolism. Dissolved CO₂ storage and leaks violate this interpretation.
Quick check · 3 questions
Geometry, coverage and scale-down

Drawn at true relative size. A deep-well plate holds a narrow column per well, a shake flask a shallow layer with a large surface, and a stirred vessel a deep pool gassed from below.
- Deep-well plate: narrow liquid column in each well
- 250 mL shake flask: shallow layer with a large gas surface
- 2 L stirred vessel: deep liquid
- Sparger and impeller supply gas from below
This empirical sulphite correlation uses shaking frequency n, liquid fill Vₗ, orbital diameter d₀ and flask diameter d. Use ratios with consistent units inside the original geometry range. Wetting, baffles, viscosity and out-of-phase motion can change transfer in real medium. Validate in the actual broth.
| Format | What can go wrong |
|---|---|
| Shake flask | A wall spot needs liquid contact. Gas exposure biases the reading toward the headspace. Verify sensor coverage over an orbit. Update liquid and headspace volumes after each sample. Humidity, condensate, leakage and evaporation bias gas balances. |
| Deep-well plate | Optical spots give pH and DO in small volumes. Edge evaporation changes concentration. Bubbles scatter light and cross-well droplets can contaminate neighbours. Repeated sampling changes fill. |
| Stirred microbioreactor | Independent mixing and feeding can reproduce fed-batch conditions with little material. Gradients, sampling disturbance and probe response can differ from a larger vessel. |
A batch screen may favour a clone with rapid sugar uptake. The same clone can rank lower when feed limits sugar flux. Oxygen limitation can mask medium effects. Match measured kLa as a first scale-down criterion, then check DO trajectories, mixing time, gas composition, feed gradients and the biological endpoint.
Quick check · 2 questions
Sources and further reading
- W. Klöckner and J. Büchs, “Shake-Flask Bioreactors”, Comprehensive Biotechnology, 3rd ed., edited by M. Moo-Young, Elsevier, 2019, ch. 2.24.
- P. Žnidaršič-Plazl and I. Plazl, “Microbioreactors”, Comprehensive Biotechnology, 3rd ed., edited by M. Moo-Young, Elsevier, 2019, ch. 2.29.
- T. Anderlei et al., “HEK293 Cell-Based Bioprocess Development at Bench Scale by Means of Online Monitoring in Shake Flasks (RAMOS and SFR)”, Methods in Molecular Biology, vol. 2095, Humana, 2020, ch. 6.
Review and apply
Revisit the opening decision
Module flashcards
Confidence by objective
Choose low to bring the related cards forward for review.
BioProcess Tools · Module summary
Small-scale and noninvasive monitoring
Key relationships
- OTR = −ΔpO₂ Vgas / (R T Δt Vliquid), during a valid closed phase.
- For a dry, isothermal closed gas with conserved inert, ΔpCO₂ = Δptotal − ΔpO₂.
Faults and checks
| Observation | Interpretation or check |
|---|---|
| Pressure slope spans valve reopening | Place both interval ends inside one closure. |
| Wall spot reads high intermittently | Inspect wetting over the orbit. |
Numbers to remember
- For kPa and L, R = 8.314 kPa·L/(mol·K).
- T in kelvin = temperature in °C + 273.15.