Locate the instruments on a benchtop reactor. Decide how much a reading can tell you about the culture.

You will be able to

  • Choose a measurement path that meets a decision deadline.
  • Separate measurement bias from process changes.
  • Interpret response time and reporting limits.
One-minute decision

Explore the instrumented vessel

Select a sensor on the vessel or use the instrument selector. Follow its position from the sterile boundary to the measured quantity.

Quick check · 3 questions

From culture to signal

A culture consumes nutrients and, in aerobic operation, oxygen. It releases heat, carbon dioxide and other products. A change in dissolved oxygen can come from a change in demand, a change in transfer or a faulty sensor. A second measurement helps distinguish these explanations.

A cutaway glass vessel with a pH electrode and a dissolved oxygen probe through the headplate. The pH cable runs to a transmitter showing pH 7.02 and on to a controller; the DO cable runs behind it straight to the controller, whose screen shows pH, dissolved oxygen and temperature trends.
3D figureFrom culture to signal

A pH electrode senses the broth at the point of contact. Its signal travels by cable to a transmitter and on to the controller, where it becomes a trend. The DO probe has its own input on the controller.

  1. pH electrode tip in the broth
  2. Headplate port, where the probe crosses the sterile boundary
  3. Probe cable
  4. Transmitter converting the electrode signal
  5. Controller trend display
Measurement path Example and consequence
In-line A pH electrode senses broth at the point of contact. Its response includes diffusion at the glass and junction.
On-line An automated line diverts exhaust to a gas analyser. Tubing introduces travel time. A liquid sample stream may be returned or discarded.
At-line An operator takes a sample to a glucose analyser beside the vessel. Record the sampling time and the reporting time.
Off-line Dry weight is determined away from the process. The delayed result can anchor the calibration of a faster sensor.

Access and inference

An immersed probe crosses the sterile boundary. A reader interrogating a spot through an intact wall is noninvasive, although the dye patch contacts the liquid. A removed sample may be consumed during analysis. These distinctions describe different aspects of a method.

An RTD gives temperature after a known resistance conversion. A soft sensor combines hardware signals with a model to estimate an unmeasured state, such as biomass from carbon dioxide production. Validate it against an independent biomass reference.

Process analytical technology (PAT) uses timely measurements and process understanding to support a quality decision. Select the decision first, then select the sensor.

Quick check · 2 questions

Sensor doctor: investigate a DO reading

A second probe provides an independent reference during a stable hold. Compare the two traces and choose a diagnosis and a corrective action. Module 2 contains the full six-case lab.

Quick check · 2 questions

Accuracy and precision

Bias moves the mean away from a reference. Poor precision spreads repeated readings. A tight cluster can therefore give a consistently wrong control action. Read the target, classify the instrument and reveal the measurement statistics.

Quick check · 2 questions

How quickly does a probe respond?

A first-order sensor reaches 90% of a step after t90 = ln(10) × τ, where τ is its time constant. A fast sampling clock can record a slow sensor many times before it settles. Move the time cursor to inspect that mismatch.

Quick check · 2 questions

What a specification promises

Quantity Question to ask
Range and resolution Which interval is calibrated? What is the smallest display increment? Extra displayed digits do not improve accuracy.
Drift and repeatability Does a stable reference move over hours? How wide is the spread on repeated measurements under the same conditions?
Detection and quantification Can a signal be distinguished from blank variation? Is the response precise enough to report a concentration?
Sterilisability Can this assembly survive the intended steam cycle or irradiation dose? Can it be checked in the installed state?

A screening convention places detection at the blank mean plus 3 standard deviations (σ) and quantification at the mean plus 10σ. Convert the signal via the calibration curve. Verify low-level precision and recovery before specifying an LLOQ.

Quick check · 2 questions

Sources and further reading

  • J. H. T. Luong et al., “Instrumentation and Analytical Methods”, Comprehensive Biotechnology, 3rd ed., edited by M. Moo-Young, Elsevier, 2019, ch. 2.60.
  • R. Pörtner (ed.), Biopharmaceutical Manufacturing: Progress, Trends and Challenges, Springer, 2024 print edition, ch. 2, “Recent Developments in Bioprocess Monitoring Systems”.
  • K. Polizzi, “Biosensors”, Comprehensive Biotechnology, 3rd ed., edited by M. Moo-Young, Elsevier, 2019, ch. 1.42.

Review and apply

Revisit the opening decision
Module flashcards
Confidence by objective

Choose low to bring the related cards forward for review.

BioProcess Tools · Module summary

Why monitor?

Key relationships

Faults and checks

ObservationInterpretation or check
Two probes disagreeCompare both with an independent reference at the same time.
Stable readings miss the referenceCheck calibration bias separately from repeatability.

Numbers to remember