Biomass and metabolites

Compare optical density with viable biovolume. Judge a metabolite estimate against an independent reference.

You will be able to

  • Separate particulate biomass from membrane-intact biovolume.
  • Recognise optical, dielectric and fluorescence interferences.
  • Align process spectra with reference assays.
One-minute decision

Biomass through a death phase

Compare two measurements calibrated against the same initial culture. Scrub through the run. An optical signal can remain large while membrane integrity and viable biovolume decline.

Reference What is counted
Dry weight Harvest a known volume, wash and dry to constant mass. Retained solids and salts add positive bias.
Optical density Blank against matching medium. Dilute into the verified linear interval and multiply by the dilution factor. Record path length, instrument and wavelength.
Cell count and viability Microscopy or cytometry counts particles within a defined gate. Dye exclusion measures membrane integrity. Colony formation measures culturability. Clumps and gates affect the result.
Quick check · 2 questions

Optical, dielectric and fluorescent signals

Transmission falls as particles scatter light out of the beam. Backscatter collects part of the light returning toward the source. Under Mie scattering, particle size, refractive index, angle and wavelength shape the signal. Multiple scattering at high density bends the calibration. Bubbles, dead cells and debris can all contribute.

The tip of a capacitance probe with four platinum ring electrodes on an insulating tip, standing in broth. Smooth field lines arc between the rings. Live cells close to the electrodes are tinted teal on one side and amber on the other along the field, with small charge beads at the poles; live cells further away are untinted. One deflated, wrinkled, dark cell leaks its contents and shows no charge separation.
3D figureCapacitance probe in broth

In the field between the ring electrodes, intact membranes charge and polarise. A dead cell with a broken membrane does not, so the signal tracks viable biovolume. Cells are drawn enlarged.

  1. Annular platinum electrodes
  2. Field lines in the broth
  3. Viable cell in the field, membrane polarised
  4. Dead cell, leaky membrane, no polarisation
  5. Insulating probe tip

Across the dielectric β-dispersion, intact membranes charge in a lower-frequency electric field. The increment relative to a higher frequency tracks viable biovolume. Cell size can therefore change the relationship to viable cell count. Membrane breakdown, bubbles, changing conductivity and electrode deposits can affect the signal. A dual-frequency difference reduces some background shifts.

Reduced nicotinamide cofactors emit near 460 nm when excited near 340 nm. The NAD(P)H signal depends on both biomass and redox state. A metabolic shift at constant biomass can therefore resemble growth. Verify the calibration in the intended medium.

Quick check · 2 questions

Metabolites and spectroscopic measurements

At-line analysers measure glucose, lactate or ammonium against standards. Sampling, filtration, queue time and assay duration create delay. A glucose oxidase sensor can detect peroxide electrochemically. Ascorbate or other electroactive species may add current. Check blanks, spike recovery and enzyme activity. Match each reference to the time the sample left the vessel.

A benchtop bioreactor with a stainless Raman probe through a headplate port; a small teal glow marks its window in the broth. A yellow fibre-optic cable runs from the probe head behind the vessel to a spectrometer whose screen shows a Raman spectrum on a sloping fluorescence background.
3D figureIn-line Raman probe

The probe window sits in the broth. A fibre-optic cable links the probe to the spectrometer, which displays the Raman spectrum for multivariate calibration.

  1. Probe tip in the broth; the glow below marks the window
  2. Raman probe in a headplate port
  3. Fibre-optic cable
  4. Spectrometer
  5. Raman spectrum on screen
Method Information and limitation
Near infrared (NIR) Broad, overlapping overtone and combination bands. Water and scattering complicate concentration estimates.
Mid infrared (MIR) Fundamental vibrations give chemical fingerprints. Strong water absorption can require a short optical path.
Raman Inelastically scattered light reports vibrational shifts. Background fluorescence, bubbles and low analyte signal can limit precision.
2D fluorescence An excitation–emission matrix records several fluorophores at once. Non-fluorescent glucose may be inferred only through correlated process changes.

These spectra contain more than one source of variation. Multivariate calibration uses measured reference concentrations across representative runs to associate spectral variation with an analyte.

Quick check · 2 questions
Professional

Validate a chemometric model

Split whole batches by run, lot and scale before fitting. Fit all transformations on training data. Standard normal variate (SNV) normalises each spectrum. Multiplicative scatter correction (MSC) adjusts offset and scale against a training reference. Derivatives can suppress baseline drift but amplify noise. Choose smoothing and latent factors inside grouped cross-validation.

PLS regresses reference values on latent spectral directions. RMSECV summarises errors during model selection. RMSEP uses untouched prediction batches. Examine bias, residuals across concentration, uncertainty and coverage near the control limit. PCA score distances and residuals can flag extrapolation. A component can track feed composition instead of the intended product.

The enzyme-lot failure

A fluorescence model reported a glucose prediction error of 0.19 g/L. In a held-out run, predicted glucose was about 1 g/L too high. The run used a different enzyme lot and stored medium. The authors proposed these changes as possible explanations for a shift in the indirect fluorescence–glucose relationship. Stop model-driven decisions, confirm reference alignment, then test additional lots as complete held-out batches.

Deploy Raman with an independent check

Acquire a reference at the scan timestamp. Challenge the model with bubbles, density, optical fouling and low glucose. A state estimator can combine Raman-PLS with a material balance and a reference assay. A biased spectrum and a biased model can agree. Use an independent measurement, monitor residuals and specify a fallback.

Quick check · 2 questions

Sources and further reading

  • K. Polizzi, “Biosensors”, Comprehensive Biotechnology, 3rd ed., edited by M. Moo-Young, Elsevier, 2019, ch. 1.42.
  • J. Glassey, “Multivariate Modeling for Bioreactor Monitoring and Control”, in C.-F. Mandenius (ed.), Bioreactors: Design, Operation and Novel Applications, Wiley-VCH, 2016, ch. 13.
  • C. Herwig (ed.), Hybrid Modelling and Multi-Parametric Control of Bioprocesses, MDPI, 2018, Schwab, Lauber and Hesse, chapter on fluorometric in situ monitoring of an E. coli cell factory.
  • B. Hitzmann (ed.), Bioprocess Monitoring and Control, MDPI, 2020, Feidl et al., chapter on combining a mechanistic model and Raman spectroscopy during antibody chromatography.

Review and apply

Revisit the opening decision
Module flashcards

Professional review cards

Confidence by objective

Choose low to bring the related cards forward for review.

BioProcess Tools · Module summary

Biomass and metabolites

Key relationships

Faults and checks

ObservationInterpretation or check
High OD during loss of viabilityCompare membrane integrity and debris against independent samples.
Single wavelength risesTest spectral overlap, background and an independent analyte assay.

Numbers to remember